Step1:
Powering Up System
Step2: Draping Tray
Arm
> Step3:
Set Up Fluidics
Step4: Connect Handpiece
and Prime Fluidics Set
Step5: Switching to
I/A Cannula
Step3:
Set Up Fluidics |
 |
3.1. Close
Roller Clamp on Irrigation Line. |
 |
3.2. Open
Drip Chamber air vent and spike bottle of BSS. Fill drip
chamber at least half full. |
 |
3.3. Hang
bottle of BSS on I/V hook of Roller Stand and raise it high. |
 |
3.4. Open
pump cover. With tubing to the left side, place collection
bag onto pins. Lay silicon tubing across center of pump rollers. |
 |
3.5. Hold
silicon tubing at the location of fitting and without stretching,
place onto left side of V-block and close cover. |
 |
3.6. Turn
Flow Control to “Prime” (8 o’clock position). |
 |
3.7. Grasp
ends of the irrigation silicon tubing at the location of
fittings and slide it up and down and forcefully work it
fully into Pinch Valve slot. Finally pull down to rest upper
fitting into slot to prevent tubing from kinking. |
 |
3.8. Turn
Flow Control to “Standby” (9 o’clock position). |
| |
Step4:
Connect Handpiece and Prime Fluidics Set > |